HOME
COMPANY
PRODUCT
GAS ANALYSIS
APPLICATION
NEWS
CONTACT
The principle and selection precautions of trace moisture analyzer in gas

2021-03-251637

BACK TO LIST

The principle and selection precautions of trace moisture analyzer in gas

1. Basic concepts

1.1 Humidity

In the measurement law, humidity is defined as the "quantity of the state of the object". Normally refers to the proportion of water contained in a certain substance.

Absolute humidity: The mass (g) of water vapor per unit volume (1m3) of gas.

Relative humidity: The ratio of the water vapor pressure (E) in the gas to the saturated water vapor pressure (Es) of the gas, expressed as a percentage: RH=E/Es×100%.

The humidity referred to in daily life is relative humidity.

1.2 Saturated water vapor pressure

The amount of water vapor in the gas is limited, and the state that reaches the limit is called saturation, and the water vapor pressure at this time is called the saturated water vapor pressure.

1.3 Dewpoint

Gases at different temperatures have different qualities of water vapor. A gas with a higher temperature also contains more water vapor. The saturated water vapor pressure corresponding to each temperature is constant. When a certain volume of gas is uniformly cooled under a constant pressure, the partial pressure of the gas and the moisture in the gas remains unchanged, the quality of the water vapor does not change, and the relative humidity increases, when it drops to a certain temperature Td When the relative humidity RH reaches 100% saturation, at this time, if the cooling is continued, until the moisture in the gas reaches a saturated state, part of the water vapor will condense into dew. The temperature at this time is the temperature at which the number of water molecules in the gas reaches dynamic equilibrium to determine the dew point of the gas, which is the dewpoint temperature, that is, the dewpoint temperature. If the dew point freezes below 0℃, it is the Frost Point.

The relative humidity is obtained from the dew point according to temperature and pressure to indirectly measure moisture, but it is greatly affected by pressure and temperature. Therefore, it is divided into normal pressure dew point and pressure dew point, and they can be interchanged.

1.4 How to express humidity

It is expressed as the dew point temperature value in ℃. It is expressed as the ratio of the sub-volume of water vapor in the wet gas to the sub-volume of dry gas, in ppm (v). Expressed by the mass of water vapor per unit volume of moisture, in g/m3.

Conversion between values in different representation methods. For example: the dew point value is -40℃=126.8ppm=0.09491g/m3.

1.5 Conversion table of dew point, humidity and ppm

Dew point 

ºC

Saturated vapor 

pressure mmHg

Relative humidity %

g/Kg

ppm

-100

0.0000099

0.000053

0.0081

0.013

-98

0.000015

0.00008

0.012

0.0197

-96

0.000022

0.00012

0.018

0.0289

-94

0.000033

0.00018

0.027

0.4034

-92

0.000048

0.00026

0.039

0.0632

-90

0.00007

0.00037

0.05

0.0921

-88

0.0001

0.00054

0.082

0.132

-86

0.00014

0.00075

0.11

0.184

-84

0.0002

0.00107

0.16

0.263

-82

0.00029

0.00155

0.24

0.382

-80

0.0004

0.00214

0.33

0.526

-78

0.00056

0.003

0.46

0.737

-76

0.00077

0.0041

0.63

1.01

-74

0.00105

0.00559

0.86

1.38

-72

0.00143

0.00762

1.17

1.88

-70

0.00194

0.0104

1.58

2.55

-68

0.00261

0.014

2.13

3.43

-66

0.00349

0.0187

2.84

4.59

-64

0.00464

0.0248

3.79

6.11

-62

0.00614

0.0328

5.01

8.08

-60

0.00808

0.043

6.59

10.6

-58

0.0106

0.0565

8.63

13.9

-56

0.0138

0.0735

11.3

18.2

-54

0.0178

0.0948

14.5

23.4

-52

0.023

0.123

18.8

30.3

-50

0.0295

0.157

24.1

38.8

-48

0.0378

0.202

30.9

49.7

-46

0.0481

0.257

39.3

63.3

-44

0.0609

0.325

49.7

80.8

-42

0.0768

410

62.7

101

-40

0.0966

0.516

78.9

127

-38

0.1209

0.644

98.6

159

-36

0.1507

0.804

122.9

198

-34

0.1873

1

152

246

-32

0.2318

1.24

189

305

-30

0.2859

1.52

234

376

-28

0.351

1.88

287

46.2

-26

0.43

2.3

351

566

-24

0.526

2.81

430

692

-22

0.64

3.41

523

842

-20

0.776

4.13

633

1020

-18

0.939

5

770

1240

-16

1.132

6.03

925

1490

-14

1.361

7.25

1110

1790

-12

1.632

8.69

1335

2150

-10

1.95

10.4

1596

2570

-8

2.326

12.4

1900

3060

-6

2.765

14.7

2260

3640

-4

3.28

17.5

2680

4320

-2

3.88

20.7

3170

5100

0

4.579

24.4

3640

6020

2

5.294

28.2

4330

6970

4

6.101

32.5

4990

3030

6

7.013

37.4

5730

9230

8

8.045

42.9

6580

10590

10

9.209

49.1

7530

12120

12

10.52

56.1

8600

13840

14

11.99

63.9

9800

15780

16

13.63

72.6

11140

17930

18

15.48

82.5

12650

20370

20

17.54

93.5

14330

23072.37


2. Common Moisture Analyzer Detection Principles

2.1 Electrolysis method:

The electrolytic moisture analyzer was successfully tested by Keide in 1959 and applied to low-content moisture measurement. Belongs to the absolute value of the direct measurement of moisture!

The sensor pillar is plated with parallel spiral platinum layer as the electrode,  and the hydrated phosphorus pentoxide film is coated between the platinum layer. Phosphorus pentoxide has a strong water absorption, when chlorine gas flows steadily through the sensor flow cell, where the water is absorbed to generate phosphoric acid, the reaction formula is as follows:

P2O5 + 3H2O → 2H3PO4

At the same time, between the two platinum layers to plus DC voltage, that has the electrolysis reaction, phosphoric acid is reductively decomposed into oxygen,chlorine, phosphorus pentoxide.The reaction formula is as follows:

4H3PO4→6H2+3O2+2P2O5 

When the absorption and electrolysis reach a balance, the water entering the electrolytic cell is absorbed by the phosphorus pentoxide film and electrolyzed. According to Faraday's law of electrolysis and gas law, it can be deduced that the electrolysis current of water is proportional to the water content of the gas sample. The specific calculation relationship is as follows:

       I = QPT0FU×10-4/3P0TV0

       I-electrolysis current of water, μA;

       U-water content of the gas sample μL/L (volume ratio);

       Q-gas sample flow mL/min;

       P—environmental pressure, Pa;

       T0=273.15K;

       F=96485C;

       P0=101325 Pa;

       T—the absolute temperature of the environment, K;

       V0=22.4L/mol.

image.png


The electrolysis moisture analyzer is an absolute measurement method, which is stable and does not drift. It can be used for acid gas such as chlorine, hydrogen chloride, hydrogen sulfide, hydrogen fluoride, sulfur dioxide, sulfur trioxide, or used for air, nitrogen, hydrogen, oxygen, argon, helium, neon, carbon monoxide, carbon dioxide, sulfur hexafluoride , Methane, ethane, propane, butane, natural gas and other neutral gases, but not suitable for alkaline gases that react with P2O5, such as ammonia.

The phosphorous pentoxide coating needs to be regenerated regularly, and the maintenance cost is very low, usually using phosphoric acid regeneration every 3 to 6 months. When used in high-humidity situations, the life of the phosphorus pentoxide coating will be shortened, and the regeneration cycle needs to be shortened.

2.2.1 Glass pillar sensor

Wrap the positive and negative electrode platinum wires on the outer or inner wall of the glass cylinder.

2.2.2 Ceramic pillar sensor

Based on the glass sensor, MZD have developed a ceramic sensor that can realize standardized and automation production. Positive and negative platinum layers are plated on ceramic flat plates. The contact area between the platinum layer and the sample gas is larger, the reaction is faster, and the measurement is more stable and reliable.


2.2 Laser method:

Pass the light of a certain wavelength range through the sample gas, and compare the detected spectrum with the reference spectrum, so that some relative changes or absorption of certain wavelengths of the spectrum will be measured. The absorption is proportional to the moisture content in the sample gas. The gas composition needs to be clean, the small range is 0-10ppm, the accuracy is high, and it can be used in the electronics industry; the large range is 0-2000ppm, but the accuracy is 100ppm. Not suitable for occasions contaminated by impurity particles.


2.3 Optical fiber method:

The surface of the fiber optic humidity sensor is a laminated structure composed of silicon oxide and zirconium oxide with different reflection coefficients. Through advanced thermal curing technology, the aperture of the sensor surface is controlled to 0.3nm, and water molecules of 0.28nm can penetrate. The controller emits a beam of near-infrared light of 790~820nm, which is transmitted to the sensor through an optical fiber cable. The water molecules entering the sensor will change the reflection coefficient of the light, thereby causing a change in wavelength, and the amount of change corresponds to the moisture content of the medium. The proportional relationship. By measuring the wavelength of the received light, the dew point and moisture content of the medium can be obtained. High precision, maintenance-free, very stable, can measure corrosive media containing H2S, HCL, etc.; the transmission fiber is easy to break and needs to be protected, and it is not suitable for occasions contaminated by impurity particles.


2.4 Dew point method

The relative humidity is obtained from the dew point according to temperature and pressure to indirectly measure moisture, but it is greatly affected by pressure and temperature. Therefore, it is divided into normal pressure dew point and pressure dew point, and they can be interchanged.


2.4.1 Chilled Mirror Method

When the gas to be measured enters the detection chamber under constant pressure, it passes the mirror at a certain flow rate. When the temperature of the mirror is higher than the dew point temperature of the gas, most of the incident dry mirror light is reflected to the photodetector, and is controlled by an amplifier and cooling. Reduce the temperature of the mirror surface. When the temperature of the mirror surface is lower than the dew point temperature of the gas, the incident light is diffusely reflected on the mirror surface due to condensation or frost, and the signal of the photodetector is weakened. Through the amplifier, the heating side is controlled to reduce the cooling capacity. Finally, the mirror temperature is automatically balanced on the dew point of the sample gas, and the temperature of the mirror at this time is measured by the thermal resistance element as the dew point temperature.

2.4.1.1 The measurement is accurate, stable and no drift, the highest accuracy instrument can reach ±0.3℃.

2.4.1.2 The price is higher, the requirements for operators are higher, and maintenance is required.

2.4.1.3 Sometimes there is supercooled water in the range of -20℃~0℃, and the cold mirror method cannot distinguish between the ice deposited on the mirror surface and the supercooled water, because the water vapor pressure is different from the ice vapor pressure, when the mirror is cooled Below 0ºC, ice may form immediately or depending on the cleanliness of the mirror and gas, even before it becomes ice at minus 44 Cº, it may be super-cooled water. This means that the readings of the instrument may have errors of up to 4 degrees Celsius. Therefore, the accuracy claimed by the cold mirror manufacturer (±0.2ºC) is absurd! Therefore, be especially careful to distinguish between cold water and frost.

2.4.1.4 The sample gas needs to be clean, sensitive to pollutants, and not resistant to corrosion. It is not suitable for occasions with impurity particles pollution and corrosiveness.

2.4.1.4.1 When there are water-soluble substances on the mirror surface, the equilibrium vapor pressure at this time will be lower than the saturated vapor pressure of pure water without impurities, which will make the measured dew point temperature higher than the actual sample gas dew point temperature, which is called Lawu Seoul effect.

2.4.1.4.2 When there are non-water-soluble impurities on the mirror surface, the background scattering level of the mirror surface will be changed, resulting in zero point shift.

2.4.1.5 When condensation is formed on the metal surface, due to the effect of surface tension, the saturated water vapor pressure on the curved surface of the lens will increase, which will make the measured dew point temperature lower than the actual sample gas dew point temperature, which is called the Kelvin effect . In the actual measurement of dew point, the Kelvin effect and the Raoul effect are just opposite, and they will cancel each other out. But the influence of the Raoult effect is much greater than the Kelvin effect, because water-soluble impurities inevitably exist in the mirror and the gas to be measured more or less, and the impurities in the gas may sometimes be related to the water-insoluble substances on the mirror. A chemical reaction or a photochemical reaction occurs, and it is converted into a soluble substance.

2.4.1.6 In the actual working process of the instrument, the mirror surface does not start to condense uniformly. The condensation layer always appears on a certain area of the mirror surface first. It is probably due to the flaws such as scratches on the mirror surface. The remaining substances are not easy to remove. On the other hand, the edges and corners of defects act as "nucleus" and accelerate the condensation process.

2.4.2 Capacitance method

Dew point meter for measuring relative humidity. Usually the measurement range is -60~20 °C, and the lowest can be measured at −80 °C.

Dew point meter for measuring absolute humidity. Usually the measurement range is -100~20 °C, and the lowest can measure -120 °C.

The capacitive dew point method is an indirect measurement of moisture. It is cheap and not resistant to corrosion. Operation at higher temperatures or certain gases will cause drift. It must be calibrated regularly to eliminate aging, lag and pollution, and post-maintenance costs are high! When the moisture content of the sample gas will be less than 10ppm for a long time, a dew point meter that measures absolute humidity should be selected.

2.4.2.1 Film capacitance sensor

The thin film sensor is depositedon two conductive electrodes, about 1~10μm thick polyamine salt or cellulose acetate polymer film, when the film absorbs or loses water, it will change the dielectric constant between the two electrodes, thereby measuring Dew point. Some high temperature resistant thermosetting polymer sensors can continuously measure high humidity and moisture at 200°C.

The thin film sensor measures the dew point by measuring the relative humidity, so its measurement range is wide, and the dew point measurement range can reach -50℃~100℃. Since all relative humidity sensors are sensitive to temperature, if they are calibrated at one temperature, they will cause errors when used at different temperatures. Although the polymer sensor is less dependent on temperature, the error is smaller when the operating temperature is different from the calibration temperature. The accuracy can reach ±1%RH in a very narrow range, and it can reach ±3%RH in a wide range of temperature and humidity. But for occasions with higher accuracy requirements, temperature compensation is required.

The thin film sensor has fast response, wide temperature and humidity measurement range, good linearity, good stability and repeatability, low temperature coefficient, low cost, and needs regular calibration. It is usually used in medium and high temperature and humidity occasions. It is not recommended to use in low humidity below 10ppm occasion.

2.4.2.2 Alumina Capacitance Sensor

The capacitance sensor is composed of an aluminum plate separated by a layer of aluminum oxide (AL2O3) and a layer of gold to form two positive and negative plates. When gas passes through the sensor, the moisture absorption layer of aluminum oxide causes a change in capacitance. At this time, the impedance of the capacitor is proportional to the partial pressure of water vapor. The partial pressure of water vapor can be obtained by measuring the impedance or capacitance of the capacitor, and the dew point value can be obtained by conversion.

The accuracy of capacitance sensor in the middle and high humidity range is generally ±1~±2℃; in the low humidity range, its accuracy is generally ±2~±3℃. This type of sensor is not resistant to corrosion. Although it does not react with hydrocarbon gases, CO, CO2, and chlorofluorocarbon gases, it will have different drifts for different gases.


2.5 Quartz Crystal Method (QCM)

After the surface coating of the quartz crystal with hygroscopic coating absorbs a certain amount of moisture, the resonant frequency of the quartz crystal decreases due to the increase in mass, so the change in the resonant frequency of the quartz crystal reflects the change in moisture. Quartz crystal vibration moisture analyzer can be used to measure the moisture content (ppbV and ppmV) of industrial pure gas, and its measurement is reliable, fast and accurate.


3. Measuring method of trace moisture in different kinds of gases

No

Industry

Product

Measuring

Measuring method

Note

1

Industial gas

liquid oxygen

moisture

Weight


2

SF6

moisture

Weight, Electrolysis, Dewpoint

Weight prciple is standard

3

Ethylene

moisture

Karl-Fischer, Hygrometer(Piezoelectric, electrolytic and capacitive)


4

Propene

moisture

Karl-Fischer Coulomb, Hygrometer(Piezoelectric and capacitive)


5

Butadiene

moisture

Karl-Fischer


6

Difluorochloromethane

moisture

Karl-Fischer


7

Isobutane

moisture

Karl-Fischer, Electrolysis

Karl-Fischer principle is standard

8

Isobutane

moisture

Karl-Fischer


9

Industial gas or 

Semiconductor gas

Nitrogen Trioxide

moisture

Piezoelectric, Electrolysis

Electrolysisis standard

10

SF6

moisture

Electrolysis, Dewpoint


11

Chlorine

moisture

Electrolysis

Electrolysis is standard

12

Silane

moisture

Electrolysis

Electrolysis is standard

13

Phosphine

moisture

spectroscopy

pectroscopy principle is standard

14

Picarro

moisture

spectroscopy

Spectroscopy principle is standard

15

Ammonia

moisture

Dewpoint

Dewpoint is standard

16

Hydrogen oxide

moisture

Dewpoint


17

Hydrogen

moisture

spectroscopy

Spectroscopy principle is standard

18

Oxygen

moisture

spectroscopy

Spectroscopy principle is standard

19

Nitrogen

moisture

spectroscopy

Spectroscopy principle is standard

20

Helium

moisture

Electrolysis

Electrolysis is standard

21

Argon

moisture

Electrolysis

Electrolysis is standard

22

Medical gas

Oxygen

moisture

Dew-Point, Electrolysis

Electrolysis is standard

23

Ammonia

moisture

Weight, Electrolysis, Dew-Point

Weight principle is standard

24

Argon

moisture

Electrolysis, Dewpoint

Dewpoint is standard

25

High purity Argon

moisture

ElectrolysisDewpoint

Dewpoint is standard

26

purity  Ammonia

moisture

Electrolysis


27

High purity Ammonia

moisture

Electrolysis


28

Neon

moisture

ElectrolysisDewpoint


29

Methane

moisture

Electrolysis


30

Oxygen

moisture

Electrolysis

Electrolysis  is standard

31

Ammonia

moisture

ElectrolysisDewpoint

Dewpoint is standard

32

Hydrogen

moisture

Dewpoint


33

Xenon

moisture

ElectrolysisDewpoint

Dewpoint is standard

34

Krypton

moisture

ElectrolysisDewpoint

Dewpoint is standard

35

Natural gas

moisture

Dewpoint


36

Dimethyl ether

moisture

Karl-Fischer Coulomb, Coulometric, Karl-Fischer Volumetric

Karl-Fischer Coulomb(Flash injection sample) is standard




This website uses cookies. Further information can be found in our Privacy policy.

OK Reject